Sciweavers

ISCAS
2006
IEEE

Electro-chemical multi-channel integrated neural interface technologies

13 years 10 months ago
Electro-chemical multi-channel integrated neural interface technologies
Abstract— We present a comparative review of two multichannel integrated neural interface technologies. The first integrated neural interface prototype performs simultaneous currentmode acquisition of 16 independent channels of redox currents ranging five orders of magnitude in dynamic range over four scales down to hundreds of picoamperes. The second neural interface acquires neural field potentials in microvolts to millivolts range on a 16x16-electrode microarray in voltage mode. Each microsystem features programmable gain amplifiers, tunable band filters, configurable sample-and-hold circuits, and is ready for external analog-to-digital conversion. The currentmode and voltage-mode neural interface prototypes have been experimentally validated in chemical and electrical neural activity monitoring respectively. Side-by-side quantitative comparison of the two neural interface technologies is given.
J. N. Y. Aziz, Roman Genov
Added 12 Jun 2010
Updated 12 Jun 2010
Type Conference
Year 2006
Where ISCAS
Authors J. N. Y. Aziz, Roman Genov
Comments (0)