Sciweavers

ADHOCNETS
2009
Springer

Oxybuoy: Constructing a Real-Time Inexpensive Hypoxia Monitoring Platform

13 years 11 months ago
Oxybuoy: Constructing a Real-Time Inexpensive Hypoxia Monitoring Platform
We present a low-cost sensor buoy designed for unattended dissolved oxygen measurement in aquatic environment. We describe the buoy’s architectural design and three experiments that we carried out to demonstrate its viability: a laboratory test in a water tank with controlled oxygen level, a lake deployment and an extensive electric power consumption test. Our experiments indicate the viability of such an approach to dissolved oxygen monitoring and promise further advances in hypoxia studies. Department of Computer Science Kent State University Technical Report: TR-KSU-CS-2009-01
Rizal Mohd Nor, Mikhail Nesterenko, Peter Lavrenty
Added 25 May 2010
Updated 25 May 2010
Type Conference
Year 2009
Where ADHOCNETS
Authors Rizal Mohd Nor, Mikhail Nesterenko, Peter Lavrentyev
Comments (0)