Sciweavers

MICCAI
2004
Springer

Rigid Registration of Freehand 3D Ultrasound and CT-Scan Kidney Images

14 years 5 months ago
Rigid Registration of Freehand 3D Ultrasound and CT-Scan Kidney Images
This paper presents a method to register a preoperative CT volume to a sparse set of intraoperative US slices. In the context of percutaneous renal puncture, the aim is to transfer a planning information to an intraoperative coordinate system. The spatial position of the US slices is measured by localizing a calibrated probe. Our method consists in optimizing a rigid 6 degree of freedom (DOF) transform by evaluating at each step the similarity between the set of US images and the CT volume. The images have been preprocessed in order to increase the relationship between CT and US pixels. Correlation Ratio turned out to be the most accurate and appropriate similarity measure to be used in a Powell-Brent minimization scheme. Results are compared to a standard rigid point-to-point registration involving segmentation, and discussed.
Antoine Leroy, Pierre Mozer, Yohan Payan, Jocelyne
Added 15 Nov 2009
Updated 15 Nov 2009
Type Conference
Year 2004
Where MICCAI
Authors Antoine Leroy, Pierre Mozer, Yohan Payan, Jocelyne Troccaz
Comments (0)